Corticosterone Kit CLIA
Aperçu rapide pour Corticosterone Kit CLIA (ABIN5653333)
Antigène
Voir toutes Corticosterone (CORT) Kits CLIAReactivité
Méthode de détection
Type de méthode
Gamme de detection
Application
Type d'échantillon
-
-
Seuil minimal de détection
- 781 pg/mL
-
Fonction
- Corticosterone (Cort) CLIA Kit for detecting Cort in serum, plasma, urine
-
Analytical Method
- Quantitative
-
Specificité
- This assay has high sensitivity and excellent specificity for detection of Corticosterone (Cort). No significant cross-reactivity or interference between Corticosterone (Cort) and analogues was observed.
-
Sensibilité
- 275 pg/mL
-
Ingrédients
-
- Pre-coated, ready to use 96-well strip plate
- Standar
- Detection Reagent A
- Detection Reagent B
- Substrate A
- Substrate B
- Assay Diluent A
- Assay Diluent B
- Wash Buffer (30x concentrate)
- Standard Diluent
- Plate Sealer
- Instruction Manual
-
Matériel non inclus
-
- Microplate Reader with 450 nm filter
- Single or multi-channel pipettes with high precision and disposable tips
- Microcentrifuge tubes
- Deionized or distilled watter
- Absorbent paper for blotting the microplate
- Container for wash solution
- 0.01 M (or 1x) Phosphate Buffered Saline (PBS), ph 7.0-7.2
-
-
-
-
Indications d'application
- Optimal working dilution should be determined by the investigator.
-
Commentaires
-
This product is tested and assembled before shipping to ensure that it meets its quality specifications. All kits are tested to confirm that they fall within their defined Inter- and Intra- assay coefficient of variation.
-
Volume d'échantillon
- 50 μL
-
Durée du test
- 2 h
-
Plaque
- Pre-coated
-
Protocole
- The microplate provided in this kit has been pre-coated with a monoclonal antibody specific to Corticosterone (Cort). A competitive inhibition reaction is launched between biotin labeled Corticosterone (Cort) and unlabeled Corticosterone (Cort) (Standards or samples) with the pre-coated antibody specific to Corticosterone (Cort). After incubation the unbound conjugate is washed off. Next, avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. The amount of bound HRP conjugate is reverse proportional to the concentration of Corticosterone (Cort) in the sample. Then the mixture of substrate A and B is added to generate glow light emission kinetics. Upon plate development, the intensity of the emitted light is reverse proportional to the Corticosterone (Cort) level in the sample or standard.
-
Précision du teste
-
Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Corticosterone (Cort) were tested 20 times on one plate, respectively
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Corticosterone (Cort) were tested on 3 different plates, 8 replicates in each plate. CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12% -
Restrictions
- For Research Use only
-
-
-
Conseil sur la manipulation
- Do not allow to contact skin or eyes. Calibrators, controls and specimen samples should be assayed in duplicate. Once the procedure has been started, all steps should be completed without interruption.
-
Stock
- 4 °C,-20 °C
-
Stockage commentaire
- -20°C. Bring all reagents to room temperature before beginning test. The kit may be stored at 4°C for immediate use within two days upon arrival. Reseal any unused strips with desiccant pack. Minimize freeze/thaw cycles.
-
Date de péremption
- 12 months
-
-
- Corticosterone (CORT)
-
Autre désignation
- Corticosterone
-
Classe de substances
- Hormone
-
Sujet
- Gene Aliases: Corticosrone
Antigène Voir toutes Corticosterone (CORT) Kits CLIA
-