HSPH1 Kit ELISA
Aperçu rapide pour HSPH1 Kit ELISA (ABIN8056632)
Antigène
Voir toutes HSPH1 Kits ELISAReactivité
Méthode de détection
Type de méthode
Gamme de detection
Application
Type d'échantillon
Classe de qualité
-
-
Seuil minimal de détection
- 0.156 ng/mL
-
Fonction
- High Sensitive ELISA Kit for Heat Shock 105kDa/110kDa Protein 1 (HSPH1)
-
Analytical Method
- Quantitative
-
Specificité
-
This assay has high sensitivity and excellent specificity for detection of High Sensitive Heat Shock 105 kDa/110 kDa Protein 1 (HSPH1).
No significant cross-reactivity or interference between High Sensitive Heat Shock 105 kDa/110 kDa Protein 1 (HSPH1) and analogues was observed. -
Sensibilité
- < 0.06 ng/mL
-
Ingrédients
-
- Pre-coated, 96-well strip plate
- Plate sealer
- Standard
- Standard Diluent
- Detection Reagent A
- Assay Diluent A
- Detection Reagent B
- Assay Diluent B
- TMB Substrate
- Stop Solution
- Wash Buffer (30X)
- Instruction manual
-
Matériel non inclus
-
- Microplate reader with 450 ± 10nm filter.
- Precision single or multi-channel pipettes and disposable tips.
- Eppendorf Tubes for diluting samples.
- Deionized or distilled water.
- Absorbent paper for blotting the microtiter plate.
- Container for Wash Solution
-
-
-
-
Volume d'échantillon
- 100 μL
-
Durée du test
- 3 h
-
Plaque
- Pre-coated
-
Protocole
-
1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 2 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.
-
Précision du teste
-
Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level High Sensitive Heat Shock 105kDa/110kDa Protein 1 (HSPH1) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level High Sensitive Heat Shock 105kDa/110kDa Protein 1 (HSPH1) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%
-
Restrictions
- For Research Use only
-
-
-
Conseil sur la manipulation
-
The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end. -
Stock
- 4 °C,-20 °C
-
Stockage commentaire
-
- For unopened kit: All the reagents should be kept according to the labels on vials. The Standard, Detection Reagent A, Detection Reagent B and the 96-well strip plate should be stored at -20°C upon receipt while the others should be at 4°C.
- For opened kit: When the kit is opened, the remaining reagents still need to be stored according to the above storage condition. Besides, please return the unused wells to the foil pouch containing the desiccant pack, and reseal along entire edge of zip-seal.
Note: It is highly recommended to use the remaining reagents within 1 month provided this is within the expiration date of the kit. - For ELISA kit, 1 day storage at 37°C can be considered as 2 months at 4°C, which means 3 days at 37°C equaling 6 months at 4°C.
-
Date de péremption
- 6 months
-
-
- HSPH1 (Heat Shock 105kDa/110kDa Protein 1 (HSPH1))
-
Autre désignation
- Heat Shock 105kDa/110kDa Protein 1
-
Sujet
- HSP105A, HSP105B, NY-CO-25, Heat shock protein 105 kDa, Heat shock 110 kDa protein
-
UniProt
- Q92598
Antigène Voir toutes HSPH1 Kits ELISA
-