SAA Kit ELISA

Le kit ELISA SAA (ABIN956172) Singe est un kit ELISA Colorimetric conçu pour quantifier Singe SAA.
N° du produit ABIN956172
1.017,00 €
Plus frais de livraison 40,00 € et TVA
96 tests
Destination: France
Envoi sous 15 à 22 jours ouvrables

Aperçu rapide pour SAA Kit ELISA (ABIN956172)

Antigène

Voir toutes SAA Kits ELISA
SAA (Serum Amyloid A (SAA))

Reactivité

  • 13
  • 12
  • 7
  • 5
  • 5
  • 4
  • 4
  • 3
  • 3
  • 3
  • 2
  • 2
  • 1
Singe

Méthode de détection

Colorimetric

Type de méthode

Sandwich ELISA

Application

ELISA

Type d'échantillon

Serum
  • Analytical Method

    Quantitative

    Attributs du produit

    The Monkey Serum Amyloid A ELISA is based on a solid phase enzyme-linked immunosorbent assay (ELISA). The assay uses two different affinity purified peptide-specific polyclonal anti-monkey SAA antibodies, one for solid phase immobilization and the other, conjugated to horseradish peroxidase (HRP), for detection. The test sample is first denatured by heating serum for 1 hour at 60°C. The denaturing step dissociates SAA from interfering factors. Subsequently, the denatured sample is diluted and incubated in the microtiter wells together with the HRP conjugate for one hour. This results in SAA molecules being sandwiched between the immobilization and detection antibodies. The wells are then washed to remove unbound HRP-labeled antibodies and TMB Reagent is added and incubated for 20 minutes at room temperature during which a blue color develops. Color development is stopped by the addition of Stop Solution, changing the color to yellow, and optical density is measured spectrophotometrically at 450 nm. The concentration of SAA is proportional to the optical density of the test sample.

    Ingrédients

    Anti-monkey SAA antibody coated microtiter plate with 96 wells (provided as 12 x 8-well strips)
    HRP Conjugate Reagent, 11 mL
    Monkey SAA Calibrator (0.20 mL, lyophilized) Note: The SAA Calibrator used in this kit is of non-monkey origin. It behaves identically to old-world monkey SAA. The use of a non-monkey calibrator allows export of the kit without the requirement for CITES documentation.
    Wash Buffer (20X), 50 mL
    Sample Diluent, 50 mL
    TMB Reagent (One-Step), 11 mL
    Stop Solution (1N HCl), 11 mL.

    Matériel non inclus

    Precision pipettes and tips
    Distilled or de-ionized water
    Polypropylene tubes
    Vortex mixer
    60°C water bath
    Absorbent paper or paper towels
    Micro-Plate incubator/shaker mixing speed of ~150 rpm
    Plate reader capable of measuring absorbance at 450 nm
    Graph paper (PC graphing software is optional)
  • Plaque

    Pre-coated

    Préparation de l'échantillon

    Denaturation
    1. Dispense 100 µL of serum into a polypropylene microcentrifuge tube.
    2. Repeat this procedure for each sample to be tested. 3. Incubate the samples at 60°C in a water bath for one hour. Dilution 1. After denaturation, dilute 1.0 µL of denatured sample with 249 µL of sample diluent.
    2. Repeat this procedure for each sample to be tested. This procedure gives a 250-fold dilution of the original sample and presents SAA in a form that is recognizable by the antibodies used in the kit.
    CALIBRATOR PREPARATION
    The calibrator vial contains lyophilized heat-treated SAA of known concentration (the calibrator must not be incubated at 60°C).
    1. Reconstitute the Monkey SAA Calibrator by addition of 200 µL of de-ionized or distilled water. Mix gently several times over a period of 5 minutes. (The reconstituted calibrator remains stable for at least 7 days at 4°C but should be aliquoted and frozen at or below -20°C after reconstitution if use beyond this time is intended).
    2. Label 6 polypropylene tubes as 25, 12.5, 6.25, 3.125, 1.56 and 0.78 ng/mL.
    3. Into the tube labeled 25 ng/mL, pipette 472.4 µL of calibrator diluent. Then add 27.62 µL of SAA calibrator and mix gently. This provides the working 25 ng/mL calibrator.
    4. Dispense 250 µL of calibrator diluent into the tubes labeled 12.5, 6.25, 3.125, 1.56 and 0.78 ng/mL.
    5. Pipette 250 µL of the 25 ng/mL calibrator into the tube labeled 12.5 ng/mL and mix. This provides the working 12.5 ng/mL calibrator.
    6. Similarly prepare the 6.25, 3.125, 1.56 and 0.78 ng/mL calibrators by serial dilution.

    Procédure de l'essai

    1. Secure the desired number of coated wells in the holder.
      2. Dispense 100 µL of calibrators and denatured/diluted samples into the wells (we recommend that calibrators and samples be tested in duplicate).
      3. Add 100 µL of HRP conjugate reagent into each well.
      4. Incubate on an orbital micro-plate shaker at 150 rpm at room temperature (18-25°C) for one hour.
      5. Wash and empty the microtiter wells 5 times with 1X wash solution using a plate washer (400 µL/well). The entire wash procedure should be performed as quickly as possible.
      6. Strike the wells sharply onto adsorbent paper or paper towels to remove all residual wash solution.
      7. Dispense 100 µL of TMB Reagent into each well.
      8. Gently mix on an orbital micro-plate shaker at 150 rpm at room temperature (18-25°C) for 20 minutes.
      9. Stop the reaction by adding 100 µL of Stop Solution to each well.
      10. Gently mix. It is important to make sure that all the blue color changes to yellow.
      11. Read the optical density at 450 nm with a microtiter plate reader within 5 minutes.

    Calcul des résultats

    1. Calculate the average absorbance values (A450) for each set of reference calibrators, and samples.
      2. Construct a calibration curve by plotting the mean absorbance obtained from each reference calibrator against its concentration in ng/mL on linear graph paper, with absorbance values on the vertical or Y-axis and concentration on the horizontal or X-axis.
      3. Using the mean absorbance value for each sample, determine the corresponding concentration of SAA in ng/mL from the calibration curve.
      4. Multiply the derived concentration by the dilution factor to determine the actual concentration of SAA in the serum sample.
      5. If available, PC graphing software should be used for the above steps. We find that a good fit of the data is obtained with either a two site binding equation or a second order polynomial equation.
      6. If the A450 values of samples fall outside the calibration curve when tested at a dilution of 250 fold, samples should be diluted appropriately and re-tested.

    Restrictions

    For Research Use only
  • Stock

    -20 °C

    Stockage commentaire

    Upon receiving the kit please store the SAA calibrator in a freezer at or below -20°C. The remaining components of the kit should be stored in a refrigerator at 4°C. It is important that the microtiter plate should be kept in a sealed bag with desiccant to minimize exposure to damp air. The kit will remain stable until the expiration date provided that the components are stored as described above.

    Date de péremption

    The expiry date is stated on the label.
  • Antigène Voir toutes SAA Kits ELISA

    SAA (Serum Amyloid A (SAA))

    Sujet

    SAA is an acute phase serum protein that can be elevated approximately 70-fold in monkeys. As is the case in humans, measurement of SAA provides an excellent biomarker of inflammation and disease.