Tel:
+49 (0)241 95 163 153
Fax:
+49 (0)241 95 163 155
E-Mail:
orders@anticorps-enligne.fr

Recoverin Kit ELISA

RCVRN Reactivité: Humain Colorimetric Sandwich ELISA 0.313-20 ng/mL Cell Culture Cells, Plasma, Serum, Tissue Lysate
N° du produit ABIN5564586
  • Antigène Voir toutes Recoverin (RCVRN) Kits ELISA
    Recoverin (RCVRN)
    Reactivité
    • 3
    • 1
    Humain
    Méthode de détection
    Colorimetric
    Type de méthode
    Sandwich ELISA
    Gamme de detection
    0.313-20 ng/mL
    Seuil minimal de détection
    0.313 ng/mL
    Application
    ELISA
    Fonction
    The AssayMax™ Human Recoverin ELISA (Enzyme-Linked Immunosorbent Assay) kit is designed for detection of human recoverin in plasma, serum, tissue extracts, and cell culture samples. This assay employs a quantitative sandwich enzyme immunoassay technique that measures human recoverin in approximately 5 hours. A polyclonal antibody specific for human recoverin has been pre-coated onto a 96-well microplate with removable strips. Recoverin in standards and samples is sandwiched by the immobilized polyclonal antibody and biotinylated polyclonal antibody specific for human recoverin, which is recognized by a streptavidin-peroxidase conjugate. All unbound material is washed away and a peroxidase enzyme substrate is added. The color development is stopped and the intensity of the color is measured.
    Marque
    AssayMax™
    Type d'échantillon
    Cell Culture Cells, Plasma, Serum, Tissue Lysate
    Analytical Method
    Quantitative
    Ingrédients
    Human Recoverin Microplate: A 96-well polystyrene microplate (12 strips of 8 wells) coated with a polyclonal antibody against human recoverin. Sealing Tapes: Each kit contains 3 precut, pressure sensitive sealing tapes that can be cut to fit the format of the individual assay. Human Recoverin Standard: Human recoverin in a buffered protein base (80 ng, lyophilized). Biotinylated Human Recoverin Antibody (50x): A 50-fold concentrated biotinylated polyclonal antibody against human recoverin (140 l). MIX Diluent Concentrate (10x): A 10-fold concentrated buffered protein base (30 ml). Wash Buffer Concentrate (20x): A 20-fold concentrated buffered surfactant (30 ml, 2 bottles). Streptavidin-Peroxidase Conjugate (SP Conjugate): A 100-fold concentrate (80 l). Chromogen Substrate: A ready-to-use stabilized peroxidase chromogen substrate tetramethylbenzidine (8 ml). Stop Solution: A 0.5 N hydrochloric acid to stop the chromogen substrate reaction (12 ml).
    Matériel non inclus
    Microplate reader capable of measuring absorbance at 405 nm. Pipettes (1-20 µL, 20-200 µL, and multiple channel). Deionized or distilled reagent grade water Incubator (37 °C)
    Featured
    Discover our best selling RCVRN Kit ELISA
    Top Product
    Discover our top product RCVRN Kit ELISA
  • Plaque
    Pre-coated
    Protocole
    • Step 1. Add 50 μL of Standard or Sample per well. Incubate 2 hours.
    • Step 2. Wash, then add 50 μL of Biotinylated Antibody per well. Incubate 2 hours.
    • Step 3. Wash, then add 50 μL of SP Conjugate per well. Incubate 30 minutes.
    • Step 4. Wash, then add 50 μL of Chromogen Substrate per well. Incubate 15 minutes.
    • Step 5. Add 50 μL of Stop Solution per well. Read at 450 nm immediately.
    Préparation des réactifs

    Freshly dilute all reagents and bring all reagents to room temperature before use. MIX Diluent Concentrate (10x): If crystals have formed in the concentrate, mix gently until the crystals have completely dissolved. Dilute the MIX Diluent Concentrate 10-fold with reagent grade water. Store for up to 30 days at 2-8 °C. Human Recoverin Standard: Reconstitute the 80 ng of Human Recoverin Standard with 4 mL of MIX Diluent to generate a 20 ng/mL standard stock solution. Allow the standard to sit for 10 minutes with gentle agitation prior to making dilutions. Prepare duplicate or triplicate standard points by serially diluting from the standard stock solution (20 ng/mL) 2-fold with MIX Diluent to produce 10, 5, 2.5, 1.25, 0.625, and 0.313 ng/mL solutions. MIX Diluent serves as the zero standard (0 ng/mL). Any remaining stock solution should be frozen at -20 °C and used within 30 days. Avoid repeated freeze-thaw cycles. 4 Standard Point Dilution [Human Recoverin] (ng/mL) P1 1 part Standard (20 ng/mL) 20.0 P2 1 part P1 + 1 part MIX Diluent 10.0 P3 1 part P2 + 1 part MIX Diluent 5.0 P4 1 part P3 + 1 part MIX Diluent 2.5 P5 1 part P4 + 1 part MIX Diluent 1.25 P6 1 part P5 + 1 part MIX Diluent 0.625 P7 1 part P6 + 1 part MIX Diluent 0.313 P8 MIX Diluent 0.0 Biotinylated Human Recoverin Antibody (50x): Spin down the antibody briefly and dilute the desired amount of the antibody 50-fold with MIX Diluent. The undiluted antibody should be stored at -20 °C. Wash Buffer Concentrate (20x): If crystals have formed in the concentrate, mix gently until the crystals have completely dissolved. Dilute the Wash Buffer Concentrate 20-fold with reagent grade water. SP Conjugate (100x): Spin down the SP Conjugate briefly and dilute the desired amount of the conjugate 100-fold with MIX Diluent. The undiluted conjugate should be stored at -20 °C.

    Prélèvement de l'échantillon
    Plasma: Collect plasma using one-tenth volume of 0.1 M sodium citrate as an anticoagulant. Centrifuge samples at 3000 x g for 10 minutes and collect plasma. Samples can be stored at -20 °C or below for up to 3 months. Avoid repeated freeze-thaw cycles. Serum: Samples should be collected into a serum separator tube. After clot formation, centrifuge samples at 3000 x g for 10 minutes and remove serum. Samples can be stored at -20 °C or below for up to 3 months. Avoid repeated freeze-thaw cycles. Tissue: Extract tissue samples with 50 mM phosphate-buffered saline ( pH 7.4) containing 1 % Triton X-100. Centrifuge at 14000 x g for 20 minutes. Collect the supernatant and measure the protein concentration. User should determine optimal dilution factor. Samples can be stored at -20 ° C or below. Avoid repeated freeze-thaw cycles. Cell Culture Supernatants: Centrifuge cell culture media at 3000 x g for 10 minutes to remove debris and collect supernatants. Samples can be stored at -20 ° C or below. Avoid repeated freeze-thaw cycles.
    Procédure de l'essai

    Prepare all reagents, standard solutions, and samples as instructed. Bring all reagents to room temperature before use. The assay is performed at room temperature (20-25 °C). Remove excess microplate strips from the plate frame and return them immediately to the foil pouch with desiccants inside. Reseal the pouch securely to minimize exposure to water vapor and store in a vacuum desiccator. Add 50 l of Human Recoverin Standard or sample per well. Gently tap plate to thoroughly coat the wells. Break any bubbles that may have formed. Cover wells with a sealing tape and incubate for 2 hours. Start the timer after the last addition. Wash five times with 200 l of Wash Buffer manually. Invert the plate each time and decant the contents, hit 4-5 times on absorbent material to completely remove the liquid. If using a machine, wash six times with 300 l of Wash Buffer and then invert the plate, decanting the contents, hit 4-5 times on absorbent material to completely remove the liquid. Add 50 l of Biotinylated Human Recoverin Antibody to each well. Gently tap plate to thoroughly coat the wells. Break any bubbles that may have formed. Cover wells with a sealing tape and incubate for 2 hours. Wash the microplate as described above. 5 Add 50 l of Streptavidin-Peroxidase Conjugate to each well. Gently tap plate to thoroughly coat the wells. Break any bubbles that may have formed. Cover wells with a sealing tape and incubate for 30 minutes. Turn on the microplate reader and set up the program in advance. Wash the microplate as described above. Add 50 l of Chromogen Substrate per well. Gently tap plate to thoroughly coat the wells. Break any bubbles that may have formed. Incubate for 15 minutes or till the optimal blue color density develops. Add 50 l of Stop Solution to each well. The color will change from blue to yellow. Gently tap plate to ensure thorough mixing. Break any bubbles that may have formed. Read the absorbance on a microplate reader at a wavelength of 450 nm immediately. If wavelength correction is available, subtract readings at 570 nm from those at 450 nm to correct optical imperfections. Otherwise, read the plate at 450 nm only. Please note that some unstable black particles may be generated at high concentration points after stopping the reaction for about 10 minutes, which will reduce the readings.

    Calcul des résultats
    • Calculate the mean value of the duplicate or triplicate readings for each standard and sample.
    • To generate a standard curve, plot the graph using the standard concentrations on the x-axis and the corresponding mean 450 nm absorbance (OD) on the y-axis. The best-fit line can be determined by regression analysis using log-log or four-parameter logistic curve-fit.
    • Determine the unknown sample concentration from the standard curve and multiply the value by the dilution factor.
    Restrictions
    For Research Use only
  • Conseil sur la manipulation
    This product is for Research Use Only and is not intended for use in diagnostic procedures. Prepare all reagents (diluent buffer, wash buffer, standard, biotinylated antibody, and SP conjugate) as instructed, prior to running the assay. Prepare all samples prior to running the assay. The dilution factors for the samples are suggested in this insert. However, the user should determine the optimal dilution factor. Spin down the SP conjugate vial and the biotinylated antibody vial before opening and using contents. The Stop Solution is an acidic solution. The kit should not be used beyond the expiration date. 2
    Stock
    4 °C,-20 °C
    Stockage commentaire
    Upon arrival, immediately store components of the kit at recommended temperatures up to the expiration date. Store SP Conjugate and Biotinylated Antibody at -20°C. Store Microplate, Diluent Concentrate (10x), Wash Buffer, Stop Solution, and Chromogen Substrate at 2-8°C. Unused microplate wells may be returned to the foil pouch with the desiccant packs and resealed. May be stored for up to 30 days in a vacuum desiccator. Diluent (1x) may be stored for up to 30 days at 2-8°C. Store Standard at 2-8°C before reconstituting with Diluent and at -20°C after reconstituting with Diluent.
  • Antigène Voir toutes Recoverin (RCVRN) Kits ELISA
    Recoverin (RCVRN)
    Autre désignation
    Recoverin (RCVRN Produits)
    Synonymes
    RCV1 Kit ELISA, CAR Kit ELISA, S-modulin Kit ELISA, Rcv1 Kit ELISA, rcv1 Kit ELISA, rcvrn Kit ELISA, recoverin Kit ELISA, reco Kit ELISA, recoverin Kit ELISA, recoverin, gene 2 S homeolog Kit ELISA, recoverin, gene 2 L homeolog Kit ELISA, Recoverin Kit ELISA, RCVRN Kit ELISA, Rcvrn Kit ELISA, rcvrn.2.S Kit ELISA, rcvrn.2.L Kit ELISA, reco Kit ELISA
    Sujet
    Recoverin is a calcium-dependent inhibitor of rhodopsin kinase, a member of the EF-hand family of calcium-binding proteins involved in the transduction of light by vertebrate photoreceptors (1-3).
    ID gène
    5957
    UniProt
    P35243
    Pathways
    Regulation of G-Protein Coupled Receptor Protein Signaling, Phototransduction