Tel:
+49 (0)241 95 163 153
Fax:
+49 (0)241 95 163 155
E-Mail:
orders@anticorps-enligne.fr

AXL Kit ELISA

Ce kit ELISA Rat AXL est un kit ELISA Colorimetric conçu pour quantifier Rat AXL.
N° du produit ABIN5652127

Aperçu rapide pour AXL Kit ELISA (ABIN5652127)

Antigène

Voir toutes AXL Kits ELISA
AXL (AXL Receptor tyrosine Kinase (AXL))

Reactivité

  • 7
  • 5
  • 1
Rat

Méthode de détection

Colorimetric

Type de méthode

Sandwich ELISA

Gamme de detection

2.34 ng/mL - 150 ng/mL

Application

ELISA

Type d'échantillon

Plasma, Serum
  • Seuil minimal de détection

    2.34 ng/mL

    Analytical Method

    Quantitative

    Specificité

    This assay has high sensitivity and excellent specificity for detection of AXL Receptor Tyrosine Kinase (AXL). No significant cross-reactivity or interference between AXL Receptor Tyrosine Kinase (AXL) and analogues was observed.

    Sensibilité

    0.9 ng/mL
  • Commentaires

    The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5 % within the expiration date under appropriate storage condition. To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

    Durée du test

    3 h

    Plaque

    Pre-coated

    Protocole

    The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to AXL Receptor Tyrosine Kinase (AXL). Standards or samples are then added to the appropriate microtiter plate wells with a biotin-conjugated antibody specific to AXL Receptor Tyrosine Kinase (AXL). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain AXL Receptor Tyrosine Kinase (AXL), biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of AXL Receptor Tyrosine Kinase (AXL) in the samples is then determined by comparing the O.D. of the samples to the standard curve.

    Précision du teste

    Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level AXL Receptor Tyrosine Kinase (AXL) were tested 20 times on one plate, respectively
    Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level AXL Receptor Tyrosine Kinase (AXL) were tested on 3 different plates, 8 replicates in each plate. CV(%) = SD/meanX100
    Intra-Assay: CV<10%
    Inter-Assay: CV<12%

    Restrictions

    For Research Use only
  • Conseil sur la manipulation

    The Stop Solution is acidic. Do not allow to contact skin or eyes. Calibrators, controls and specimen samples should be assayed in duplicate. Once the procedure has been started, all steps should be completed without interruption.

    Stock

    4 °C,-20 °C

    Stockage commentaire

    -20°C. Bring all reagents to room temperature before beginning test. The kit may be stored at 4°C for immediate use within two days upon arrival. Reseal any unused strips with desiccant pack. Minimize freeze/thaw cycles.

    Date de péremption

    4-8 months
  • Antigène Voir toutes AXL Kits ELISA

    AXL (AXL Receptor tyrosine Kinase (AXL))

    Autre désignation

    AXL Receptor Tyrosine Kinase

    Sujet

    Gene Name: AXL Receptor Tyrosine Kinase

    Gene Aliases: UFO, JTK11, Tyrosine-protein kinase receptor UFO

    ID gène

    308444

    UniProt

    Q8VIA0

    Pathways

    Signalisation RTK, Cellular Response to Molecule of Bacterial Origin
Vous êtes ici:
Chat with us!