AGA Kit ELISA
Aperçu rapide pour AGA Kit ELISA (ABIN8068024)
Antigène
Voir toutes AGA Kits ELISAReactivité
Méthode de détection
Type de méthode
Gamme de detection
Application
Type d'échantillon
-
-
Seuil minimal de détection
- 1.37 units/mL
-
Fonction
- The Aspartylglucosaminidase (AGA) Human IgG ELISA Kit is an in vitro indirect ELISA for the semi-quantitative measurement of human IgG autoantibodies against AGA in human serum or plasma.
-
Analytical Method
- Quantitative
-
Ingrédients
-
- Size / Description
- AGA Coated 96-Well Microplate
- 96 wells coated with AGA
- Wash Buffer
- Positive Control
- Negative Serum Control
- Biotinylated Anti-Human IgG
- HRP-Streptavidin
- TMB One-Step Substrate Reagent
- Stop Solution
- Assay Diluent
- Sample Diluent
-
Matériel non inclus
-
- Microplate reader capable of measuring absorbance at 450 nm
- Precision pipettes to deliver 2 μL to 1 mL volumes
- Adjustable 1-25 mL pipettes for reagent preparation
- 100 mL and 1 liter graduated cylinders
- Absorbent paper
- Distilled or deionized water
-
-
-
-
Indications d'application
- Suggested Plasma Dilution: 500-fold
-
Plaque
- Pre-coated
-
Protocole
-
- Prepare all reagents, samples and standards as instructed.
- Add 100 μL positive control, negative control, or sample to each well. Incubate 1 hour at room temperature.
- Add 100 μL prepared Biotinylated Anti-Human IgG Antibody into respective wells. Incubate 1 hour at room temperature.
- Add 100 μL prepared HRP-Streptavidin solution to each well. Incubate 30 minutes at room temperature.
- Add 100 μL TMB One-Step Substrate Reagent to each well. Incubate 10 to 30 minutes at room temperature.
- Add 50 μL Stop Solution to each well. Read at 450 nm immediately.
-
Restrictions
- For Research Use only
-
-
-
Stock
- -20 °C
-
Stockage commentaire
- The entire kit may be stored at -20°C for up to 1 year from the date of shipment. Avoid repeated freeze-thaw cycles. For extended storage, it is recommended to store at -80°C.
-
-
- AGA (Aspartylglucosaminidase (AGA))
-
Autre désignation
- Aspartylglucosaminidase
-
Sujet
- Optional[Gene Symbols]: The Aspartylglucosaminidase (AGA) Human IgG ELISA Kit employs an indirect ELISA method. In this kit, standard 96-well plates (12 strips with 8 wells/strip) are coated with AGA protein, which combines with the corresponding antibody present in a sample and Positive Control, which is used as a calibration curve for interpretation purposes. When a biotinylated secondary anti-human IgG antibody is added, a complex forms between biotinylated secondary antibody and human antibody bound to the protein coated plate. A TMB substrate is added, and a blue color is generated. The depth of color is relative to the amount of the human anti-AGA antibody present. The Stop Solution changes the color from blue to yellow, and the intensity of the color is measured at 450 nm.
Antigène Voir toutes AGA Kits ELISA
-