Tel:
+49 (0)241 95 163 153
Fax:
+49 (0)241 95 163 155
E-Mail:
orders@anticorps-enligne.fr

Lactate Dehydrogenase A Kit ELISA

LDHA Reactivité: Rat Colorimetric Sandwich ELISA 0.31 ng/mL - 20 ng/mL Plasma, Serum
N° du produit ABIN5519280
  • Antigène Voir toutes Lactate Dehydrogenase A (LDHA) Kits ELISA
    Lactate Dehydrogenase A (LDHA)
    Reactivité
    • 4
    • 3
    • 3
    • 1
    • 1
    • 1
    Rat
    Méthode de détection
    Colorimetric
    Type de méthode
    Sandwich ELISA
    Gamme de detection
    0.31 ng/mL - 20 ng/mL
    Seuil minimal de détection
    0.31 ng/mL
    Application
    ELISA
    Fonction
    The kit is a sandwich enzyme immunoassay technique for the in vitro quantitative measurement in various sample types.
    Type d'échantillon
    Plasma, Serum
    Analytical Method
    Quantitative
    Sensibilité
    0.19 ng/mL
    Ingrédients
    • Pre-coated, ready to use 96-well strip plate, flat buttom
    • Plate sealer for 96 wells
    • Reference Standard
    • Reference Standard & Sample Diluent
    • Biotinylated Detection Antibody (100 x concentrate)
    • HRP Conjugate (100 x concentrate)
    • Biotinylated Detection Antibody Diluent
    • HRP Conjugate Diluent
    • Substrate Reagent
    • Stop Solution
    • Wash Buffer (25 x concentrate)
    • Instruction manual
    Featured
    Discover our best selling LDHA Kit ELISA
    Top Product
    Discover our top product LDHA Kit ELISA
  • Volume d'échantillon
    100 µL
    Durée du test
    1 - 4.5 h
    Plaque
    Pre-coated
    Protocole
    1. Add 100 µL standard or sample to each well. Incubate for 90 min at 37 °C.
    2. Remove the liquid. Add 100 µL Biotinylated Detection Antibody. Incubate for 1 hour at 37 °C.
    3. Aspirate and wash 3 times.
    4. Add 100 µL HRP Conjugate. Incubate for 30 min at 37 °C.
    5. Aspirate and wash 5 times.
    6. Add 90 µL Substrate Reagent. Incubate for 15 min at 37 °C.
    7. Add 50 µL Stop Solution. Read at 450 nm immediately.
    8. Calculation of results.
    Préparation des réactifs
    1. Bring all reagents to room temperature (18~25 °C) before use. Follow the Microplate reader manual for set-up and preheat it for 15 min before OD measurement.
    2. Wash Buffer: Dilute 30 mL of Concentrated Wash Buffer with 720 mL of deionized or distilled water to prepare 750 mL of Wash Buffer.Note: if crystals have formed in the concentrate, warm it in a 40 °C water bath and mix it gently until the crystals have completely dissolved
    3. Standard working solution: Centrifuge the standard at 10,000xg for 1 min. Add 1.0 mL of Reference Standard &Sample Diluent, let it stand for 10 min and invert it gently several times. After it dissolves fully, mix it thoroughly with a pipette. This reconstitution produces a working solution of 20 ng/mL. Then make serial dilutions as needed. The recommended dilution gradient is as follows: 20, 10, 5, 2.5, 1.25, 0.63, 0.31, 0 ng/mL. Dilution method: Take 7 EP tubes, add 500 μLof Reference Standard & Sample Diluent to each tube. Pipette 500 μLof the 20 ng/mL working solution to the first tube and mix up to produce a 10 ng/mL working solution. Pipette 500 μLof the solution from the former tube into the latter one according to these steps. The illustration below is for reference. Note: the last tube is regarded as a blank. Don't pipette solution into it from the former tube.
    4. Biotinylated Detection Antibody working solution: Calculate the required amount before the experiment (100 μL/well). In preparation, slightly more than calculated should be prepared. Centrifuge the stock tube before use, dilute the 100x Concentrated Biotinylated Detection Antibody to 1xworking solution with Biotinylated Detection Antibody Diluent.
    5. Concentrated HRP Conjugate working solution: Calculate the required amount before the experiment (100 μL/well). In preparation, slightly more than calculated should be prepared. Dilute the 100x Concentrated HRP Conjugate to 1x working solution with Concentrated HRP Conjugate Diluent.
    Restrictions
    For Research Use only
  • Stock
    4 °C/-20 °C
    Stockage commentaire
    The unopened kit can be stored at 4℃ for 1 month. If the kit is not used within 1 month, store the items separately according to the conditions since the kit is received.
  • Antigène Voir toutes Lactate Dehydrogenase A (LDHA) Kits ELISA
    Lactate Dehydrogenase A (LDHA)
    Autre désignation
    LDHA(Lactate Dehydrogenase A) (LDHA Produits)
    Synonymes
    GSD11 Kit ELISA, LDH1 Kit ELISA, LDHM Kit ELISA, LDH-M Kit ELISA, ldh-h Kit ELISA, ldha Kit ELISA, ldha-B Kit ELISA, ldhbb Kit ELISA, trg-5 Kit ELISA, Ldh1 Kit ELISA, Ldhm Kit ELISA, l7R2 Kit ELISA, gsd11 Kit ELISA, ldh-m Kit ELISA, ldh1 Kit ELISA, ldhm Kit ELISA, pig19 Kit ELISA, ldha1 Kit ELISA, ldhab Kit ELISA, LDH-A Kit ELISA, LDHC Kit ELISA, lactate dehydrogenase A Kit ELISA, lactate dehydrogenase B L homeolog Kit ELISA, lactate dehydrogenase A4 Kit ELISA, lactate dehydrogenase A L homeolog Kit ELISA, L-lactate dehydrogenase A chain Kit ELISA, LDHA Kit ELISA, ldhb.L Kit ELISA, Ldha Kit ELISA, ldha Kit ELISA, ldha.L Kit ELISA, LOC100713875 Kit ELISA
    Pathways
    L'effet Warburg
Vous êtes ici:
Support technique