CD4 Kit ELISA
Aperçu rapide pour CD4 Kit ELISA (ABIN8068296)
Antigène
Voir toutes CD4 Kits ELISAÉpitope
Reactivité
Méthode de détection
Type de méthode
Application
Type d'échantillon
-
-
Fonction
- Measures nitrated CD4 and total CD4 in human cell and tissue lysates, enabling precise assessment of CD4 nitration by ELISA. This assay can be developed for other species (mouse, rat, etc.) upon request. Contact techsupport@raybiotech.com.
-
Analytical Method
- Quantitative
-
Ingrédients
-
- Size / Qty
- Microplate Coated With Anti-Pan-CD4 Antibody
- Anti-Nitration Detection Antibody
- Anti-CD4 Detection Antibody
- HRP-Conjugated Streptavidin
- 200 μL
- Wash Buffer
- Assay Diluent
- Lysis Buffer
- TMB One-Step Substrate Reagent
- Stop Solution
-
Matériel non inclus
-
- Microplate reader capable of measuring absorbance at 450 nm
- Protease and Phosphatase inhibitors
- Precision pipettes to deliver 2 μL to 1 mL volumes
- Adjustable 1-25 mL pipettes for reagent preparation
- 100 mL and 1 liter graduated cylinders
- Log-log graph paper or computer and software for ELISA data analysis
- Absorbent paper
- Tubes to prepare sample dilutions
- Distilled or deionized water
- Shaker
-
-
-
-
Indications d'application
- Optimal working dilution should be determined by the investigator.
-
Plaque
- Pre-coated
-
Protocole
-
- Prepare all reagents and samples as instructed.
- Add 100 μL sample to each well. Incubate 2.5 hours at room temperature or overnight at 4 °C with gentle shaking.
- Add 100 μL prepared detection antibody to each well. Incubate for 1 hour at room temperature with gentle shaking.
- Add 100 μL prepared HRP-Conjugated Streptavidin solution. Incubate for 1 hour at room temperature with gentle shaking.
- Add 100 μL TMB One-Step Substrate Reagent to each well. Incubate 30 minutes at room temperature.
- Add 50 μL Stop Solution to each well. Read at 450 nm immediately.
-
Restrictions
- For Research Use only
-
-
-
Stock
- -20 °C
-
Stockage commentaire
- The entire kit may be stored at -20°C for up to 6 months from the date of shipment. Avoid repeated freeze-thaw cycles.
-
-
- CD4
-
Autre désignation
- CD4
-
Sujet
-
Gene Symbols: CD4
Background: The Nitrated-CD4 and Total-CD4 ELISA Kit is an in vitro enzyme-linked immunosorbent assay for the measurement of human nitrated ATP-citrate synthase and total ATP-citrate synthase. An anti-pan CD4 antibody has been coated onto a 96-well plate. Samples are pipetted into the wells and CD4 present in a sample is bound to the wells by the immobilized antibody. The wells are washed and biotinylated anti-nitration antibody is used to detect nitrated CD4 or biotinylated anti-CD4 antibody is used to detect total CD4 . After washing away unbound antibody, HRP-conjugated streptavidin is pipetted to the wells. The wells are again washed, a TMB substrate solution is added to the wells and color develops in proportion to the amount of nitrated CD4 or total CD4 bound. The Stop Solution changes the color from blue to yellow, and the intensity of the color is measured at 450 nm.
Protein Name (Synonyms): T-cell surface glycoprotein CD4, T-cell surface antigen T4/Leu-3
-
ID gène
- 920
-
UniProt
- P01730
-
Pathways
- TCR Signaling, Maintenance of Protein Location, CXCR4-mediated Signaling Events
Antigène Voir toutes CD4 Kits ELISA
-