HIF1A Kit ELISA

HIF1A Reactivité: Rat Colorimetric Sandwich ELISA 7.8-500 pg/mL
N° du produit ABIN955879
  • Antigène Voir toutes HIF1A Kits ELISA
    HIF1A (Hypoxia Inducible Factor 1, alpha Subunit (Basic Helix-Loop-Helix Transcription Factor) (HIF1A))
    Reactivité
    • 9
    • 4
    • 3
    • 3
    • 2
    • 2
    • 1
    • 1
    • 1
    • 1
    • 1
    • 1
    • 1
    Rat
    Méthode de détection
    Colorimetric
    Type de méthode
    Sandwich ELISA
    Gamme de detection
    7.8-500 pg/mL
    Seuil minimal de détection
    7.8 pg/mL
    Application
    ELISA
    Fonction
    This ELISA kit is a sandwich enzyme immunoassay for the in vitro quantitative measurement of Rat HIF1A in serum, plasma and other biological fluids.
    Analytical Method
    Quantitative
    Specificité
    This assay has high sensitivity and excellent specificity for detection of Rat HIF1A. No significant cross- reactivity or interference was observed.
    Sensibilité
    The minimum detectable dose of Rat HIF1 A is typically less than 3.7 pg/mL.
    The Sensitivity of this assay, or Lower Limit of Detection (LLD) was defined as the lowest protein concentration that could be differentiated from zero. It was determined by the mean O.D. value of 20 replicates of the zero calibrator plus three standard deviations.
    Attributs du produit
    The microtiter plate provided in this kit has been pre-coated with an antibody specific to HIF1A. Calibrators and samples are then added to the appropriate microtiter plate wells with a biotin-conjugated polyclonal antibody preparation specific for HIF1A. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. Then the TMB substrate solution is added to each well. Only those wells that contain HIF1A, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of a sulfuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450 nm +/- 2 nm. The concentration of HIF1A in the samples is then determined by comparing the O.D. of the samples to the calibration curve.
    Ingrédients
    Pre-coated, ready to use 96-well plate (1x)
    Calibrator (lyophilized) (2x)
    Calibrator Diluent (1 x 20 mL)
    Detection Reagent A (1 x 120 µL)
    Detection Reagent B (1 x 120 µL)
    Assay Diluent A (2X concentrate) (1 x 6 mL)
    Assay Diluent B (2X concentrate) (1 x 6 mL)
    TMB Substrate (1 x 9 mL)
    Stop Solution (1 x 6 mL)
    Wash Buffer (30X concentrate) (1 x 20 mL)
    Plate sealer for 96 wells (4x).
    Matériel non inclus
    1. Microplate reader with 450 +/- 10 nm filter.
    2. Precision single and multi-channel pipettes and disposable tips.
    3. Eppendorf Tubes for diluting samples.
    4. Deionized or distilled water.
    5. Absorbent paper for blotting the microtiter plate.
    6. Container for Wash Solution
    Featured
    Discover our best selling HIF1A Kit ELISA
    Top Product
    Discover our top product HIF1A Kit ELISA
  • Commentaires

    The calibration curve concentrations used for the ELISA's were 500, 250, 125, 62.5, 31.2, 15.6, 7.8 pg/mL.

    Plaque
    Pre-coated
    Calcul des résultats

    Average the duplicate readings for each calibrator, control, and sample and subtract the average zero calibrator optical density. Create a calibration curve by reducing the data using computer software capable of generating a four parameter logistic (4-PL) curve-fit. As an alternative, construct a calibration curve by plotting the mean absorbance for each calibrator on the x-axis against the concentration on the y-axis and draw a best fit curve through the points on the graph. The data may be linearized by plotting the log of the HIF1A concentrations versus the log of the O.D. and the best fit line can be determined by regression analysis. This procedure will produce an adequate but less precise fit of the data. It is recommended to use some related software to do this calculation. If samples have been diluted, the concentration read from the calibration curve must be multiplied by the dilution factor.

    Restrictions
    For Research Use only
  • Stock
    -20 °C
    Stockage commentaire
    All the reagents should be kept according to the labels on vials. The Calibrator, Detection Reagent A, Detection Reagent B and the 96-well strip plate should be stored at -20° C upon being received. The unused strips should be kept in a sealed bag with the desiccant provided to minimize exposure to damp air. Opened test kits will remain stable until the expiration date shown, provided it is stored as prescribed above.
    Date de péremption
    The expiry date is stated on the label.
  • Antigène Voir toutes HIF1A Kits ELISA
    HIF1A (Hypoxia Inducible Factor 1, alpha Subunit (Basic Helix-Loop-Helix Transcription Factor) (HIF1A))
    Autre désignation
    HIF1a (HIF1A Produits)
    Synonymes
    hif1a Kit ELISA, AA959795 Kit ELISA, HIF1alpha Kit ELISA, MOP1 Kit ELISA, bHLHe78 Kit ELISA, HIF-1A Kit ELISA, HIF-1alpha Kit ELISA, HIF1 Kit ELISA, HIF1-ALPHA Kit ELISA, PASD8 Kit ELISA, HIF1-alpha Kit ELISA, hif-1a Kit ELISA, hypoxia inducible factor 1 alpha Kit ELISA, hypoxia inducible factor 1, alpha subunit Kit ELISA, hypoxia inducible factor 1 alpha subunit Kit ELISA, HIF (Hypoxia Inducible Factor) homologa Kit ELISA, hif-1a Kit ELISA, Hif1a Kit ELISA, HIF1A Kit ELISA, hif1a Kit ELISA, hif-1 Kit ELISA
    Pathways
    Positive Regulation of Peptide Hormone Secretion, Regulation of Hormone Metabolic Process, Regulation of Hormone Biosynthetic Process, Cellular Response to Molecule of Bacterial Origin, Carbohydrate Homeostasis, Transition Metal Ion Homeostasis, Tube Formation, Regulation of Carbohydrate Metabolic Process, Signaling Events mediated by VEGFR1 and VEGFR2, VEGFR1 Specific Signals, L'effet Warburg
Vous êtes ici:
Support technique